A calibration-free ammonia breath sensor using a quantum cascade laser with WMS 2f/1f

Citation data:

Applied Physics B, ISSN: 0946-2171, Vol: 116, Issue: 2, Page: 371-383

Publication Year:
2014
Usage 75
Abstract Views 53
Full Text Views 14
Link-outs 8
Captures 35
Readers 25
Exports-Saves 10
Citations 20
Citation Indexes 20
Repository URL:
http://hdl.handle.net/10754/563157
DOI:
10.1007/s00340-013-5701-1
Author(s):
Owen, Kyle; Farooq, Aamir
Publisher(s):
Springer Nature; Springer Verlag
Tags:
Physics and Astronomy
article description
The amount of ammonia in exhaled breath has been linked to a variety of adverse medical conditions, including chronic kidney disease (CKD). The development of accurate, reliable breath sensors has the potential to improve medical care. Wavelength modulation spectroscopy with second harmonic normalized by the first harmonic (WMS 2f/1f) is a sensitive technique used in the development of calibration-free sensors. An ammonia gas sensor is designed and developed that uses a quantum cascade laser operating near 1,103.44 cmand a multi-pass cell with an effective path length of 76.45 m. The sensor has a 7 ppbv detection limit and 5 % total uncertainty for breath measurements. The sensor was successfully used to detect ammonia in exhaled breath and compare healthy patients to patients diagnosed with CKD. © 2013 Springer-Verlag Berlin Heidelberg.