PlumX Metrics
Embed PlumX Metrics

Monitoring of Breathing Pattern at Rest by Electrical Impedance Tomography

Archivos de Bronconeumología ((English Edition)), ISSN: 1579-2129, Vol: 43, Issue: 6, Page: 300-303
2007
  • 21
    Citations
  • 0
    Usage
  • 19
    Captures
  • 0
    Mentions
  • 0
    Social Media
Metric Options:   Counts1 Year3 Year

Metrics Details

Article Description

Electrical impedance tomography (EIT) involves the application of a small alternating current to produce a series of chest images that can be used to monitor breathing pattern. The relation between chest images and tidal volume has not been sufficiently validated. The aim of the present study was to analyze the correlation between EIT images and the volume–time signal measured with a pneumotachometer in 13 healthy volunteers.