Extended Calibration of Flame Atomic-Absorption Instruments by a Flow Injection Peak Width Method

Citation data:

Analyst, Vol: 109

Publication Year:
1984
Usage 13
Downloads 12
Abstract Views 1
Repository URL:
https://scholarworks.umass.edu/chem_faculty_pubs/1288
Author(s):
Tyson, Julian
Tags:
Atomic-absorption calibration; flow injection analysis; peak width; Chemistry
article description
The principles of two existing methods of calibration based on flow injection methodology are outlined and the mathematical basis of a method based on peak width is presented. On the basis of the single well stirred tank model, it is shown that peak width at any concentration is proportional to logarithm of a function of concentration. The relationship is verified experimentally over the range 1-1 000 p.p.m. of magnesium. The advantages and limitations of the method are discussed.