Measurement of therapeutic photon beams-induced Cerenkov radiation generated in PMMA- and PS-based plastic optical fibers

Optical Review, ISSN: 1349-9432, Vol: 23, Issue: 5, Page: 806-810
2016
  • 8
    Citations
  • 0
    Usage
  • 0
    Captures
  • 0
    Mentions
  • 0
    Social Media
Metric Options:   Counts1 Year3 Year

Metrics Details

  • Citations
    8

Article Description

In this study, we characterized Cerenkov radiation generated in polystyrene (PS)- and polymethyl methacrylate (PMMA)-based plastic optical fibers (POFs) to select an adequate optical fiber for producing Cerenkov radiation. To determine the relationship between the absorbed dose and the intensity of Cerenkov radiation, we calculated the energy depositions of photon beams and fluxes of electrons inducing Cerenkov radiation using the Monte Carlo N-Particle eXtended code. Also, intensities of Cerenkov radiation generated in PS- and PMMA-based POFs were measured as functions of dose rate and monitor unit. At last, therapeutic photon beams-induced Cerenkov radiation in PS- and PMMA-based POFs was measured according to depths of solid water phantom.