Continuous measurement of multiple inert and respiratory gas exchange in an anaesthetic breathing system by continuous indirect calorimetry

Journal of Clinical Monitoring and Computing, ISSN: 1387-1307, Vol: 23, Issue: 1, Page: 41-49
2009
  • 5
    Citations
  • 0
    Usage
  • 16
    Captures
  • 0
    Mentions
  • 0
    Social Media
Metric Options:   Counts1 Year3 Year

Metrics Details

Article Description

A method was tested which permits continuous monitoring from a breathing system of the rate of uptake of multiple gas species, such as occurs in patients during inhalational anaesthesia. The method is an indirect calorimetry technique which uses fresh gas rotameters for control, regulation and measurement of the gas flows into the system, with continuous sampling of mixed exhaust gas, and frequent automated recalibration to maintain accuracy. Its accuracy was tested in 16 patients undergoing pre-cardiopulmonary bypass coronary artery surgery, breathing mixtures of oxygen/air and sevoflurane with/without nitrous oxide, by comparison with the reverse Fick method. Overall mean bias [95% confidence interval (CI)] of rate of uptake was 17.9 [7.3 to 28.5] ml min for oxygen, 0.04 [-0.42 to 0.50] ml min for sevoflurane, 10.9 [-16.1 to 37.8] for CO, and 8.8 [-14.8 to 32.4] ml min for nitrous oxide where present. The method proved to be accurate and precise, and allows continuous monitoring of exchange of multiple gases using standard gas analysis devices. © Springer Science+Business Media, LLC 2009.