PlumX Metrics
Embed PlumX Metrics

Evaluation of an automated synthesis method for [F]Fluoromisonidazole using a Scintomics GRP® module

Journal of Radioanalytical and Nuclear Chemistry, ISSN: 1588-2780, Vol: 333, Issue: 11, Page: 5889-5898
2024
  • 0
    Citations
  • 0
    Usage
  • 7
    Captures
  • 0
    Mentions
  • 0
    Social Media
Metric Options:   Counts1 Year3 Year

Metrics Details

Article Description

There is limited information on utilizing commercially supplied [F]Fluoride from an off-site cyclotron for the synthesis of radiopharmaceuticals. This study explored the production of the PET hypoxia marker [F]FMISO, using the Scintomics® GRP module and distantly produced [F]Fluoride, which lacks published data. A radiochemical yield of 27.4 ± 3.6% (n = 5) and high radiochemical purity were achieved in synthesis time of 48 min. The [F]FMISO met all Ph. Eur. standards, demonstrating the consistency and reliability of the module in generating a clinical-grade radiopharmaceutical using off-site produced [F]Fluoride, but available patient doses were severely limited.