Evaluation of a Gafchromic EBT2 film dosimetry system for radiotherapy quality assurance
Australasian Physical and Engineering Sciences in Medicine, ISSN: 0158-9938, Vol: 34, Issue: 2, Page: 251-260
2011
- 97Citations
- 66Captures
Metric Options: CountsSelecting the 1-year or 3-year option will change the metrics count to percentiles, illustrating how an article or review compares to other articles or reviews within the selected time period in the same journal. Selecting the 1-year option compares the metrics against other articles/reviews that were also published in the same calendar year. Selecting the 3-year option compares the metrics against other articles/reviews that were also published in the same calendar year plus the two years prior.
Example: if you select the 1-year option for an article published in 2019 and a metric category shows 90%, that means that the article or review is performing better than 90% of the other articles/reviews published in that journal in 2019. If you select the 3-year option for the same article published in 2019 and the metric category shows 90%, that means that the article or review is performing better than 90% of the other articles/reviews published in that journal in 2019, 2018 and 2017.
Citation Benchmarking is provided by Scopus and SciVal and is different from the metrics context provided by PlumX Metrics.
Example: if you select the 1-year option for an article published in 2019 and a metric category shows 90%, that means that the article or review is performing better than 90% of the other articles/reviews published in that journal in 2019. If you select the 3-year option for the same article published in 2019 and the metric category shows 90%, that means that the article or review is performing better than 90% of the other articles/reviews published in that journal in 2019, 2018 and 2017.
Citation Benchmarking is provided by Scopus and SciVal and is different from the metrics context provided by PlumX Metrics.
Metrics Details
- Citations97
- Citation Indexes97
- 97
- CrossRef62
- Captures66
- Readers66
- 66
Article Description
This study examines the dosimetric accuracy of Gafchromic EBT2 model radiochromic film for use in radiotherapy quality assurance. In this study, film was scanned using an Epson Perfection V700 flatbed scanner in transmission mode at 75 DPI with the subsequent analysis performed using the red and blue colour channels and ImageJ software. Results of this study suggest that the conversion of film optical density to measured dose should, at present, utilise red channel data only, without application of a blue channel correction to the data. For the batch of film examined here, film uniformity and reproducibility appear to have improved compared with published results using older batches. The orientation of the film on the scanner and the side of the film facing the light source were found to have substantial effects on results. Based on the results of this study, it is possible to recommend the use of EBT2 film in routine quality assurance testing for radiotherapy, in situations where a dose uncertainty of up to 2.8% is acceptable. © 2011 Australasian College of Physical Scientists and Engineers in Medicine.
Bibliographic Details
http://www.scopus.com/inward/record.url?partnerID=HzOxMe3b&scp=80051700455&origin=inward; http://dx.doi.org/10.1007/s13246-011-0072-6; http://www.ncbi.nlm.nih.gov/pubmed/21465275; http://link.springer.com/10.1007/s13246-011-0072-6; http://www.springerlink.com/index/10.1007/s13246-011-0072-6; http://www.springerlink.com/index/pdf/10.1007/s13246-011-0072-6; https://dx.doi.org/10.1007/s13246-011-0072-6; https://link.springer.com/article/10.1007/s13246-011-0072-6