PlumX Metrics
Embed PlumX Metrics

Secondary radiation doses of intensity-modulated radiotherapy and proton beam therapy in patients with lung and liver cancer

Radiotherapy and Oncology, ISSN: 0167-8140, Vol: 98, Issue: 3, Page: 335-339
2011
  • 21
    Citations
  • 0
    Usage
  • 39
    Captures
  • 0
    Mentions
  • 0
    Social Media
Metric Options:   Counts1 Year3 Year

Metrics Details

Article Description

To compare the secondary radiation doses following intensity-modulated radiotherapy (IMRT) and proton beam therapy (PBT) in patients with lung and liver cancer. IMRT and PBT were planned for three lung cancer and three liver cancer patients. The treatment beams were delivered to phantoms and the corresponding secondary doses during irradiation were measured at various points 20–50 cm from the beam isocenter using ion chamber and CR-39 detectors for IMRT and PBT, respectively. The secondary dose per Gy (i.e., a treatment dose of 1 Gy) from PBT for lung and liver cancer, measured 20–50 cm from the isocenter, ranged from 0.17 to 0.086 mGy. The secondary dose per Gy from IMRT, however, ranged between 5.8 and 1.0 mGy, indicating that PBT is associated with a smaller dose of secondary radiation than IMRT. The internal neutron dose per Gy from PBT for lung and liver cancer, 20–50 cm from the isocenter, ranged from 0.03 to 0.008 mGy. The secondary dose from PBT is less than or compatible to the secondary dose from conventional IMRT. The internal neutron dose generated by the interaction between protons and body material is generally much less than the external neutron dose from the treatment head.