PlumX Metrics
Embed PlumX Metrics

Electrocatalytic oxidation and flow detection analysis of semicarbazide at based IrO x chemically modified electrodes

Sensors and Actuators B: Chemical, ISSN: 0925-4005, Vol: 209, Page: 25-31
2015
  • 18
    Citations
  • 0
    Usage
  • 15
    Captures
  • 0
    Mentions
  • 0
    Social Media
Metric Options:   Counts1 Year3 Year

Metrics Details

  • Citations
    18
    • Citation Indexes
      18
  • Captures
    15

Article Description

An IrO x modified electrode was characterized and proposed as an amperometric sensor for the detection of semicarbazide using cyclic voltammetry, chronoamperometry and amperometry techniques. The modified electrode was characterized as sensing probe in both acid and alkaline medium toward the electrocatalytic oxidation and detection of semicarbazide. A liquid chromatography approach based on the strong cation exchange mode (LC-SCX column) was used in acid conditions for the separation and analysis of semicarbazide in a synthetic mixture. Operating under a constant applied potential of 1.1 V vs. Ag/AgCl and using HClO 4 as carrier electrolyte, the limit of detection (LOD) is attested at level of 0.1 μM (i.e., 7.5 μg L −1 ), while an excellent precision and temporal stability of the amperometric signal, was observed. The linear ranges spanning over three order of magnitude of the detection limit with a correlation coefficients higher than 0.995.

Provide Feedback

Have ideas for a new metric? Would you like to see something else here?Let us know