Simple and effective in situ sample illumination for electron paramagnetic resonance

Chemical Communications, ISSN: 1364-548X, Vol: 60, Issue: 8, Page: 1012-1015
2024
  • 0
    Citations
  • 0
    Usage
  • 7
    Captures
  • 0
    Mentions
  • 0
    Social Media
Metric Options:   Counts1 Year3 Year

Metrics Details

Article Description

Illumination into an electron paramagnetic resonance (EPR) spectrometer is commonly carried out through the optical window, perpendicular to the sample and magnetic field. Here we show that significant improvements can be obtained by using the walls of the EPR tube as a light guide, with the light scattered only around the sample-containing area.

Bibliographic Details

Woodward, Adam W; Bramham, Jack E; Brookfield, Adam; Golovanov, Alexander P; Bowen, Alice M

Royal Society of Chemistry (RSC)