Article Description

The factors limiting the resolution of the traditional correlation deep level transient spectroscopy (DLTS) are revealed and analyzed. It is shown that all weighting functions proposed to date provided nonsymmetrical rate windows, effectively filtering out slow transients, but were mediocre filters for fast transients. It is argued that it was the response to fast transients which actually limited the resolution of the correlation DLTS. The resolution limit of previously published weighting functions is determined. It is shown that the limitations are inherent in the earlier approach to the devising of weighting functions rather than in the correlation procedure itself. It can be overcome using new weighting functions based on the Gaver-Stehfest algorithm for the inverse Laplace transformation. © 1997 American Institute of Physics.