Ultra-high-bandwidth polarization interferometry and optimal quadratic phase detection

Review of Scientific Instruments, ISSN: 1089-7623, Vol: 90, Issue: 8, Page: 083503
2019
  • 0
    Citations
  • 0
    Usage
  • 3
    Captures
  • 0
    Mentions
  • 0
    Social Media
Metric Options:   Counts1 Year3 Year

Metrics Details

Article Description

A novel homodyne interferometer and analysis method are described which use orthogonal polarization components to measure large rapid changes in interferometric phase, in quadrature, in the presence of strong time-dependent attenuation of the scene beam. This approach overcomes the major sources of error associated with homodyne interferometry (sensitivity nulls, ambiguity in the direction of phase change when passing through a sensitivity null, and intolerance to beam power variations) while maintaining its intrinsic simplicity and speed, enabling extremely high-bandwidth, high-dynamic range measurements limited only by available detector technology. Using this technique, electron density in a magnetized plasma shock was measured with unprecedented bandwidth and resolution, revealing short-timescale features not previously observed.