Effects of Charité artificial disc on the implanted and adjacent spinal segments mechanics using a hybrid testing protocol
Spine, ISSN: 0362-2436, Vol: 30, Issue: 24, Page: 2755-2764
2005
- 199Citations
- 101Captures
Metric Options: CountsSelecting the 1-year or 3-year option will change the metrics count to percentiles, illustrating how an article or review compares to other articles or reviews within the selected time period in the same journal. Selecting the 1-year option compares the metrics against other articles/reviews that were also published in the same calendar year. Selecting the 3-year option compares the metrics against other articles/reviews that were also published in the same calendar year plus the two years prior.
Example: if you select the 1-year option for an article published in 2019 and a metric category shows 90%, that means that the article or review is performing better than 90% of the other articles/reviews published in that journal in 2019. If you select the 3-year option for the same article published in 2019 and the metric category shows 90%, that means that the article or review is performing better than 90% of the other articles/reviews published in that journal in 2019, 2018 and 2017.
Citation Benchmarking is provided by Scopus and SciVal and is different from the metrics context provided by PlumX Metrics.
Example: if you select the 1-year option for an article published in 2019 and a metric category shows 90%, that means that the article or review is performing better than 90% of the other articles/reviews published in that journal in 2019. If you select the 3-year option for the same article published in 2019 and the metric category shows 90%, that means that the article or review is performing better than 90% of the other articles/reviews published in that journal in 2019, 2018 and 2017.
Citation Benchmarking is provided by Scopus and SciVal and is different from the metrics context provided by PlumX Metrics.
Metrics Details
- Citations199
- Citation Indexes199
- 199
- CrossRef155
- Captures101
- Readers101
- 101
Article Description
Study Design. Finite element model of L3-S1 segment and confirmatory cadaveric testing were used to investigate the biomechanical effects of a mobile core type artificial disc (Charité artificial disc; DePuy Spine, Raynham, MA) on the lumbar spine. Objective. To determine the effects of the Charité artificial disc across the implanted and adjacent segments. Summary of Background Data. Biomechanical studies of artificial discs that quantify parameters, like the load sharing and stresses, are sparse in the literature, especially for mobile-type core artificial disc designs. In addition, there is no standard protocol for studying the adjacent segmental effects of such implants. Methods. Human osteo-ligamentous spines (L1-S1) were tested before and after L5-S1 Charité artificial disc placement. The data were used to validate further an intact 3-dimensional (3-D) nonlinear L3-S1 finite element model. The model was subjected to 400-N axial compression and 10.6 Nm of flexion/extension pure moments (load control) or pure moments that produced the overall rotation of the L3-S1 Charité model equal to the intact case (hybrid approach). Resultant motion, load, and stress parameters were analyzed at the experimental and adjacent levels. Results. Finite element model validation was achieved only with the load-controlled experiments. The hybrid approach, believed to be more clinically relevant, revealed that Charité artificial disc leads to motion increases inflexion (19%) and extension (44%) at the L5-S1 level. At the instrumented level, the decrease in the facet loads was less than atthe adjacent levels; the corresponding decrease being 26% at L3-L4, 25% at L4-L5, and 13.4% at L5-S1 when compared to the intact. Intradiscal pressure changes in the L4-L5 and L3-L4 segments were minimal. Shear stresses at the Charité artificial disc-L5 endplate interface were higher than those at S1 interface. However, in the load control mode, the increase in facet loads in extension was approximately 14%, as compared to the intact case. Conclusions. The hybrid testing protocol is advocated because it better reproduces clinical observations in terms of motion following surgery, using pure moments. Using this approach, we found that the Charité artificial disc placement slightly increases motion at the implanted level, with a resultant increase in facet loading when compared to the adjacent segments, while the motions and loads decrease at the adjacent levels. However, in the load control mode that we believe is not that clinically relevant, there was a large increase in motion and a corresponding increase in facet loads, as compared to the intact. ©2005, Lippincott Williams & Wilkins, Inc.
Bibliographic Details
http://www.scopus.com/inward/record.url?partnerID=HzOxMe3b&scp=29444456951&origin=inward; http://dx.doi.org/10.1097/01.brs.0000195897.17277.67; http://www.ncbi.nlm.nih.gov/pubmed/16371899; http://journals.lww.com/00007632-200512150-00009; https://dx.doi.org/10.1097/01.brs.0000195897.17277.67; https://journals.lww.com/spinejournal/Abstract/2005/12150/Effects_of_Charit__Artificial_Disc_on_the.9.aspx